High reproducibility of two-dimensional liquid chromatography using pH-driven fractionation with a pressure-resistant electrode. - Inserm - Institut national de la santé et de la recherche médicale Access content directly
Journal Articles Journal of Chromatography B - Analytical Technologies in the Biomedical and Life Sciences Year : 2008

High reproducibility of two-dimensional liquid chromatography using pH-driven fractionation with a pressure-resistant electrode.

Frédéric Pont

Abstract

Automated two-dimensional liquid chromatography using the PF2D system from Beckman Coulter provides a fractionation platform well suited for differential proteomic studies. To date, the reliability and reproducibility of PF2D has not been accurately tested. Here, we used an optimized software and a pressure-resistant pH electrode, allowing a precise and reproducible control of the pH limits for each fraction during PF2D. We tested the reliability of this improved system by performing several rounds of fractionation using the same protein extract. Three UV maps were generated, leading to 54 chromatograms and more than 3000 protein peaks. Using semi-automated software for peak-to-peak comparison between 2D-LC chromatograms, we demonstrate that the peak concordance is very high. The rates of concordance were higher in the second dimension repeatability tests, indicating that the limiting factors for 2D-LC reproducibility rely on the pI fractionation and sample preparation steps. The reproducibility between maps was closely related to pH curves similarities, further stressing the need of careful pH adjustment and precise electrode calibration.
Fichier principal
Vignette du fichier
170308-PF2D-FPONT-JCB.pdf (370.56 Ko) Télécharger le fichier
inserm-00369925_edited.pdf (332.01 Ko) Télécharger le fichier
Origin : Files produced by the author(s)
Origin : Files produced by the author(s)
Loading...

Dates and versions

inserm-00369925 , version 1 (23-03-2009)

Identifiers

Cite

Elsa Suberbielle, Daniel Gonzales-Dunia, Frédéric Pont. High reproducibility of two-dimensional liquid chromatography using pH-driven fractionation with a pressure-resistant electrode.: High reproducibility of two-dimensional liquid chromatography using pH-driven fractionation with a pressure-resistant electrode.. Journal of Chromatography B - Analytical Technologies in the Biomedical and Life Sciences, 2008, 871 (1), pp.125-9. ⟨10.1016/j.jchromb.2008.06.043⟩. ⟨inserm-00369925⟩
96 View
488 Download

Altmetric

Share

Gmail Facebook X LinkedIn More