Multi-modal and multi-scale clinical retinal imaging system with pupil and retinal tracking - Inserm - Institut national de la santé et de la recherche médicale Access content directly
Journal Articles Scientific Reports Year : 2022

Multi-modal and multi-scale clinical retinal imaging system with pupil and retinal tracking

Nicolas Lefaudeux
  • Function : Author
Marine Durand
  • Function : Author
Xavier Levecq
  • Function : Author
Nicolas Chateau
  • Function : Author

Abstract

Abstract We present a compact multi-modal and multi-scale retinal imaging instrument with an angiographic functional extension for clinical use. The system integrates scanning laser ophthalmoscopy (SLO), optical coherence tomography (OCT) and OCT angiography (OCTA) imaging modalities and provides multi-scale fields of view. For high resolution, and high lateral resolution in particular, cellular imaging correction of aberrations by adaptive optics (AO) is employed. The entire instrument has a compact design and the scanning head is mounted on motorized translation stages that enable 3D self-alignment with respect to the subject’s eye by tracking the pupil position. Retinal tracking, based on the information provided by SLO, is incorporated in the instrument to compensate for retinal motion during OCT imaging. The imaging capabilities of the multi-modal and multi-scale instrument were tested by imaging healthy volunteers and patients.
Fichier principal
Vignette du fichier
s41598-022-13631-1.pdf (3.65 Mo) Télécharger le fichier
Origin : Publisher files allowed on an open archive

Dates and versions

inserm-03726478 , version 1 (18-07-2022)

Identifiers

Cite

Muhammad Faizan Shirazi, Jordi Andilla, Nicolas Lefaudeux, Claudia Valdes, Florian Schwarzhans, et al.. Multi-modal and multi-scale clinical retinal imaging system with pupil and retinal tracking. Scientific Reports, 2022, 12 (1), pp.9577. ⟨10.1038/s41598-022-13631-1⟩. ⟨inserm-03726478⟩
58 View
16 Download

Altmetric

Share

Gmail Facebook Twitter LinkedIn More