Bench Assessment of Expiratory Valve Resistance of Current ICU Ventilators in Dynamic Conditions

Résumé

BACKGROUND: We hypothesized that the lack of benefit of setting a low versus a high PEEP in patients with ARDS may be due in part to differences in the dynamic behavior of the expiratory valve in ventilators. We tested this hypothesis by conducting a bench comparison of the dynamic behavior of expiratory valves on ICU ventilators currently in use. METHODS: We attached 7 ICU ventilators (C5, C6, Carescape, PB980, ServoU, V500, and V680) to the ASL 5000 lung model (passive condition with compliance 20 mL/cm H(2)O and resistance 5 cm H(2)O/L/s) and set in volume controlled mode (tidal volume 0.8 L, breathing frequency 10 breaths/min). Flow and pressure were measured just before the exhalation valve. At PEEP of 5, 10, and 15 cm H(2)O, the median instantaneous expiratory resistance, the time to valve opening, and the pressure time products above or below the values of PEEP (expressed in cm H(2)O × s) were determined. RESULTS: Median instantaneous expiratory resistance values differed between the ventilators and PEEP settings with a significant interaction: at PEEP 5 cm H(2)O, the median (interquartile range) expiratory resistance values were 3.9 (3.5-4.7), 3.0 (3.0-3.1), 20.9 (15.8-24.9), 27.4 (26.5-43.2), 13.8 (13.6-13.9), 4.4 (4.0-4.6), and 34.3 (33.7-33.8) cm H(2)O/L/s, for the C5, C6, Carescape, PB980, ServoU, V500, and V680, respectively. For all the PEEP settings, the corresponding times to valve opening were 0.080 (0.077-0.082), 0.082 (0.080-0.085), 0.110 (0.105-0.110), 0.100 (0.085-1.05), 0.072 (0.062-0.072), 0.145 (0.115-0.150), and 0.075 (0.070-0.080) s, respectively, and pressure-time products were 2.8 (2.1-7.4), 6.8 (6.7-7.3), 2.4 (2.1-2.4), 3.5 (2.7-3.6), 1.8 (1.8-2.1), 2.8 (2.7-2.9), and 5.7 (5.4-5.9) cm H(2)O × s, respectively. CONCLUSIONS: The resistance of active expiratory valves differed significantly between the 7 ICU ventilators tested.

Dates et versions

inserm-03268238 , version 1 (23-06-2021)

Identifiants

Citer

A. Pinède, M. Cour, F. Degivry, B. Louis, L. Argaud, et al.. Bench Assessment of Expiratory Valve Resistance of Current ICU Ventilators in Dynamic Conditions. Respiratory Care, 2020, 66 (4), pp.610-618. ⟨10.4187/respcare.08098⟩. ⟨inserm-03268238⟩
22 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More