Mechanical thrombectomy with the ERIC retrieval device: initial experience

Abstract : OBJECTIVE: To report our experience with the Embolus Retriever with Interlinked Cage (ERIC) stentriever for use in mechanical endovascular thrombectomy (MET). METHODS: Thirty-four consecutive patients with acute stroke (21 men and 13 women; median age 66 years) determined appropriate for MET were treated with ERIC and prospectively included over a 6-month period at three different centers. The ERIC device differs from typical stentrievers in that it is designed with a series of interlinked adjustable nitinol cages that allow for fast thrombus capture, integration, and withdrawal. The evaluated endpoints were successful revascularization (Thrombolysis in Cerebral Infarction (TICI) 2b-3) and good clinical outcomes at 3 months (modified Rankin Scale (mRS) 0-2). RESULTS: Locations of the occlusions included the middle cerebral artery (13 patients), terminal carotid artery (11 patients), basilar artery (1 patient), and tandem occlusions (9 patients). IV thrombolysis was performed in 20/34 (58.8%) patients. Median times from symptom onset to recanalization and from puncture to recanalization were 325.5 min (180-557) and 78.5 min (14-183), respectively. Used as the first-line device, ERIC achieved a successful recanalization in 20/24 (83.3%) patients. Successful recanalization was associated with lower National Institutes of Health Stroke Scale scores at 24 h (8±6.5 vs 21.5±2.1; p=0.008) and lower mRS at 3 months (2.7±2.1 vs 5.3±1.1; p=0.04). Three procedural complications and four asymptomatic hemorrhages were recorded. Good clinical outcomes at 3 months were seen in 15/31 (48.4%) patients. CONCLUSIONS: The ERIC device is an innovative stentriever allowing fast, effective, and safe MET.
Type de document :
Article dans une revue
Journal of Neurointerventional Surgery, BMJ Journals, 2017, 9 (6), pp.574 - 577. 〈10.1136/neurintsurg-2016-012379〉
Liste complète des métadonnées

https://www.hal.inserm.fr/inserm-01936717
Contributeur : Jean-Yves Gauvrit <>
Soumis le : mardi 27 novembre 2018 - 15:34:45
Dernière modification le : vendredi 15 février 2019 - 01:29:40

Identifiants

Citation

Hélène Raoult, Hocine Redjem, Romain Bourcier, Alina Gaultier-Lintia, Benjamin Daumas-Duport, et al.. Mechanical thrombectomy with the ERIC retrieval device: initial experience. Journal of Neurointerventional Surgery, BMJ Journals, 2017, 9 (6), pp.574 - 577. 〈10.1136/neurintsurg-2016-012379〉. 〈inserm-01936717〉

Partager

Métriques

Consultations de la notice

17