Communication Dans Un Congrès Année : 2015

Comparison between 3D Supersonic Shear Wave Elastography and Magnetic Resonance Elastography: a preliminary experimental study

Résumé

Ultrasound Supersonic Shear Wave Elastog-raphy (SSWE) as well as Magnetic Resonance Elastography (MRE) allow accessing the mechanical properties of human tissues. SSWE is usually performed using a 2D probe. 3D SSWE is now available but needs to be validated. We compared 3D SSWE with both 2D SSWE and MRE which is inherently 3D on a breast phantom. We found that 3D SSWE is reproducible and provides elasticity estimates comparable to those obtained with the validated 2D SSWE. We also showed that 3D SSWE and MRE exhibit quite different elasticity moduli , but they reveal similar qualitative trends in the phantom. Although no relationship could be drawn between the two modalities , this study provides a first basis for comparison and a guide for potential improvements .
Fichier principal
Vignette du fichier
ACTES-RITS2015-HAL-p142-3.pdf (578.38 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

inserm-01155015 , version 1 (26-05-2015)

Licence

Paternité

Identifiants

  • HAL Id : inserm-01155015 , version 1

Citer

Jinlong Yue, Marion Tardieu, Felicia Julea, Linda Chami, Olivier Lucidarme, et al.. Comparison between 3D Supersonic Shear Wave Elastography and Magnetic Resonance Elastography: a preliminary experimental study. Journées RITS 2015, Mar 2015, Dourdan, France. pp 142-143. ⟨inserm-01155015⟩
299 Consultations
187 Téléchargements