Denoising arterial spin labeling MRI using tissue partial volume

Jan Petr 1, * Jean-Christophe Ferré 1, 2 Jean-Yves Gauvrit 1, 2 Christian Barillot 1
* Corresponding author
1 VisAGeS - Vision, Action et Gestion d'informations en Santé
INSERM - Institut National de la Santé et de la Recherche Médicale : U746, Inria Rennes – Bretagne Atlantique , IRISA-D5 - SIGNAUX ET IMAGES NUMÉRIQUES, ROBOTIQUE
Abstract : Arterial spin labeling (ASL) is a noninvasive MRI method that uses magnetically labeled blood to measure cerebral perfusion. Spatial resolution of ASL is relatively small and as a consequence perfusion from different tissue types is mixed in each pixel. An average ratio of gray matter (GM) to white matter (WM) blood flow is 3 to 1. Disregarding the partial volume effects (PVE) can thus cause serious errors of perfusion quantification. PVE also complicates spatial filtering of ASL images as apart from noise there is a spatial signal variation due to tissue partial volume. Recently, an algorithm for correcting PVE has been published by Asllani et al. It represents the measured magnetization as a sum of different tissue magnetizations weighted by their fractional volume in a pixel. With the knowledge of the partial volume obtained from a high-resolution MRI image, it is possible to separate the individual tissue contributions by linear regression on a neighborhood of each pixel. We propose an extension of this algorithm by minimizing the total-variation of the tissue specific magnetization. This makes the algorithm more flexible to local changes in perfusion. We show that this method can be used to denoise ASL images without mixing the WM and GM signal.
Complete list of metadatas

Cited literature [2 references]  Display  Hide  Download

https://www.hal.inserm.fr/inserm-00601147
Contributor : Jan Petr <>
Submitted on : Thursday, June 16, 2011 - 7:12:56 PM
Last modification on : Tuesday, March 5, 2019 - 1:45:43 AM
Long-term archiving on : Friday, November 9, 2012 - 3:15:27 PM

File

Petr10b.pdf
Files produced by the author(s)

Identifiers

Citation

Jan Petr, Jean-Christophe Ferré, Jean-Yves Gauvrit, Christian Barillot. Denoising arterial spin labeling MRI using tissue partial volume. SPIE Medical Imaging 2010: Image Processing, Feb 2010, San Diego, United States. pp.n.a, ⟨10.1117/12.844443⟩. ⟨inserm-00601147⟩

Share

Metrics

Record views

495

Files downloads

448