Ultrasonic focusing through the ribs using the DORT method. - Inserm - Institut national de la santé et de la recherche médicale Accéder directement au contenu
Article Dans Une Revue Medical Physics Année : 2009

Ultrasonic focusing through the ribs using the DORT method.

Résumé

Thermal ablation induced by high intensity focused ultrasound has produced promising clinical results to treat hepatocarcinoma and other liver tumors. However skin burns have been reported due to the high absorption of ultrasonic energy by the ribs. This study proposes a method to produce an acoustic field focusing on a chosen target while sparing the ribs using the decomposition of the time-reversal operator (DORT method). The idea is to apply an excitation weight vector to the transducers array which is orthogonal to the subspace of emissions focusing on the ribs. A linear array of transducers has been used to measure the set of singular vectors associated with a chest phantom, made of three human ribs immersed in water, and to produce the desired acoustic fields. The resulting propagating fields have been measured both in the focal plane and in the plane of the ribs using a needle hydrophone. The ratio of the energies absorbed at the focal point and on the ribs has been enhanced up to 100-fold, as demonstrated by the measured specific absorption rates.
Fichier principal
Vignette du fichier
MPH003495.pdf (781.06 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

inserm-00419916 , version 1 (25-09-2009)

Identifiants

  • HAL Id : inserm-00419916 , version 1
  • PUBMED : 19746783

Citer

Etienne Cochard, C. Prada, Jean-François Aubry, Mathias Fink. Ultrasonic focusing through the ribs using the DORT method.. Medical Physics, 2009, 36 (8), pp.3495-503. ⟨inserm-00419916⟩
389 Consultations
454 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More