Clinical evaluation of a computer-controlled pressure support mode.

Abstract : We have designed a computerized system providing closed-loop control of the level of pressure support ventilation (PSV). The system sets itself at the lowest level of PSV that maintains respiratory rate (RR), tidal volume (VT), and end-tidal CO(2) pressure (PET(CO(2))) within predetermined ranges defining acceptable ventilation (i.e., 12 < RR < 28 cycles/min, VT > 300 ml [> 250 if weight < 55 kg], and PET(CO(2)) < 55 mm Hg [< 65 mm Hg if chronic CO(2) retention]). Ten patients received computer-controlled (automatic) PSV and physician-controlled (standard) PSV, in random order, during 24 h for each mode. An estimation of occlusion pressure (P(0.1)) was recorded continuously. The average time spent with acceptable ventilation as previously defined was 66 +/- 24% of the total ventilation time with standard PSV versus 93 +/- 8% with automatic PSV (p < 0.05), whereas the level of PSV was similar during the two periods (17 +/- 4 cm H(2)O versus 19 +/- 6 cm H(2)O). The time spent with an estimated P(0.1) above 4 cm H(2)O was 34 +/- 35% of the standard PSV time versus only 11 +/- 17% of the automatic PSV time (p < 0.01). Automatic PSV increased the time spent within desired ventilation parameter ranges and apparently reduced periods of excessive workload.
Document type :
Journal articles
Complete list of metadatas

Cited literature [24 references]  Display  Hide  Download

https://www.hal.inserm.fr/inserm-00402420
Contributor : Michel Dojat <>
Submitted on : Tuesday, July 7, 2009 - 3:09:48 PM
Last modification on : Wednesday, August 19, 2020 - 11:16:27 AM
Long-term archiving on: : Tuesday, September 18, 2012 - 12:56:13 PM

Files

all1.PDF
Files produced by the author(s)

Identifiers

  • HAL Id : inserm-00402420, version 1
  • PUBMED : 10764306

Collections

Citation

Michel Dojat, Alain Harf, Dominique Touchard, François Lemaire, Laurence Brochard. Clinical evaluation of a computer-controlled pressure support mode.. American Journal of Respiratory and Critical Care Medicine, American Thoracic Society, 2000, 161 (4 Pt 1), pp.1161-6. ⟨inserm-00402420⟩

Share

Metrics

Record views

703

Files downloads

1035