An optical fiber-based gating device for prospective mouse cardiac MRI.

Raphaël Sablong 1, * Adrian Rengle 1 Anoop Ramgolam 1 Hervé Saint-Jalmes 2, 3 Olivier Beuf 1
* Auteur correspondant
1 5 - RMN et optique : De la mesure aux biomarqueurs
CREATIS - Centre de Recherche en Acquisition et Traitement de l'Image pour la Santé
Abstract : Prospective synchronization of MRI acquisitions on living organisms involves the monitoring of respiratory and heart motions. The electrocardiogram (ECG) signal is conventionally used to measure the cardiac cycle. However, in some circumstances, obtaining an uncorrupted ECG signal recorded on small animals with radio frequency (RF) pulses and gradient switching is challenging. To monitor respiratory motion, an air cushion associated with a pressure sensor is commonly used but the system suffers from bulkiness. For many applications, the physiological gating information can also be derived from an MR navigated signal. However, a compact device that can simultaneously provide respiratory and cardiac information, for both prospective gating and physiological monitoring, is desirable. This is particularly valid since small volume coils or dedicated cardiac RF coil arrays placed directly against the chest wall are required to maximize measurement sensitivity. An optic-based device designed to synchronize MRI acquisitions on small animal's respiratory and heart motion was developed using a transmit-receive pair of optical fibers. The suitability of the developed device was assessed on mice ( n = 10) and was based on two sets of experiments with dual cardiac and respiratory synchronization. Images acquired with prospective triggering using the optical-based signal, ECG, and the pressure sensor during the same experiment were compared between themselves in the first set. The second set compared prospective technique using optical-based device and ECG to a retrospective technique. The optical signal that was correlated to both respiratory and heart motion was totally unaffected by radiofrequency pulses or currents induced by the magnetic field gradients used for imaging. Mice heart MR images depict low-visible motion artifacts with all sensors or techniques used. No significant SNR differences were found between each series of image. Full fiber-optic-based signal derived from heart and respiratory motion was suitable for prospective triggering of heart MR imaging. The fiber optic device performed similarly to the ECG and air pressure sensors, while providing an advantage for imaging with dedicated cardiac array coils by reducing bulk. It can be an attractive alternative for small animal MRI in difficult environments such as limited space and strong gradient switching.
Type de document :
Article dans une revue
IEEE Transactions on Biomedical Engineering, Institute of Electrical and Electronics Engineers, 2014, 61 (1), pp.162-70. 〈10.1109/TBME.2013.2278712〉
Liste complète des métadonnées

Littérature citée [17 références]  Voir  Masquer  Télécharger

http://www.hal.inserm.fr/inserm-00921893
Contributeur : Lotfi Senhadji <>
Soumis le : lundi 7 juillet 2014 - 17:34:05
Dernière modification le : mercredi 16 mai 2018 - 11:23:40
Document(s) archivé(s) le : mardi 7 octobre 2014 - 10:37:13

Fichier

TBME_Optical_Device_SABLONG_R....
Fichiers produits par l'(les) auteur(s)

Identifiants

Citation

Raphaël Sablong, Adrian Rengle, Anoop Ramgolam, Hervé Saint-Jalmes, Olivier Beuf. An optical fiber-based gating device for prospective mouse cardiac MRI.. IEEE Transactions on Biomedical Engineering, Institute of Electrical and Electronics Engineers, 2014, 61 (1), pp.162-70. 〈10.1109/TBME.2013.2278712〉. 〈inserm-00921893〉

Partager

Métriques

Consultations de la notice

863

Téléchargements de fichiers

249