2191-219X-3-78 2191-219X Original research <p>Kinetic analysis of [<sup>11</sup>C]befloxatone in the human brain, a selective radioligand to image monoamine oxidase A</p> Zanotti-FregonaraPaolopaolo.zanotti-fregonara@chu-bordeaux.fr LeroyClaireclaire.leroy@cea.fr RoumenovDimitrid.roumenov@neuf.fr TrichardChristianchristiantrichard@free.fr MartinotJean-Lucjean-luc.martinot@cea.fr BottlaenderMichelmichel.bottlaender@cea.fr

University of Bordeaux, CNRS, INCIA, UMR 5287, Talence 33400, France

CEA, DSV, I2BM, Service Hospitalier Frederic Joliot, Orsay 91401, France

INSERM, CEA, ParisSud University, U.1000, Orsay 91400, France

CEA, DSV, I2BM, Neurospin, Gif-sur-Yvette 91191, France

EJNMMI Research 2191-219X 2013 3 1 78 http://www.ejnmmires.com/content/3/1/78 10.1186/2191-219X-3-7824274579
13920132110201325112013 2013Zanotti-Fregonara et al.; licensee Springer.This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. [11C]Befloxatone Monoamine oxidase A PET Spectral analysis

Abstract

Background

[11C]Befloxatone measures the density of the enzyme monoamine oxidase A (MAO-A) in the brain. MAO-A is responsible for the degradation of different neurotransmitters and is implicated in several neurologic and psychiatric illnesses. This study sought to estimate the distribution volume (V T) values of [11C]befloxatone in humans using an arterial input function.

Methods

Seven healthy volunteers were imaged with positron emission tomography (PET) after [11C]befloxatone injection. Kinetic analysis was performed using an arterial input function in association with compartmental modeling and with the Logan plot, multilinear analysis (MA1), and standard spectral analysis (SA) at both the regional and voxel level. Arterialized venous samples were drawn as an alternative and less invasive input function.

Results

An unconstrained two-compartment model reliably quantified V T values in large brain regions. A constrained model did not significantly improve V T identifiability. Similar V T results were obtained using SA; however, the Logan plot and MA1 slightly underestimated V T values (about -10%). At the voxel level, SA showed a very small bias (+2%) compared to compartmental modeling, Logan severely underestimated V T values, and voxel-wise images obtained with MA1 were too noisy to be reliably quantified. Arterialized venous blood samples did not provide a satisfactory alternative input function as the Logan-V T regional values were not comparable to those obtained with arterial sampling in all subjects.

Conclusions

Binding of [11C]befloxatone to MAO-A can be quantified using an arterial input function and a two-compartment model or, in parametric images, with SA.

Background

Monoamine oxidase (MAO) is a mitochondrial enzyme responsible for the degradation of several neurotransmitters, such as dopamine, serotonin, and noradrenaline. Two MAO isoforms - products of different genes - have been identified in mammals: enzyme monoamine oxidase A (MAO-A), which is primarily found in cathecolaminergic neurons, and MAO-B, which is mainly present in glial cells and serotonergic neurons 1 . MAO-A is widespread in the human brain; the highest concentrations are found in the thalamus and the occipital cortex, intermediate values are found in the basal ganglia, and lower concentrations are seen in the frontal cortex and cerebellum 1 2 . MAO-A dysfunction has been implicated in several neurologic illnesses such as Parkinson's disease, Alzheimer's disease, and Huntington's disease, as well as in psychiatric diseases such as major depressive disorder 3 . Notably, several studies have linked MAO-A enzyme activity and gene expression to tobacco addiction 2 4 5 6 7 .

With some exceptions 6 , studies on MAO-A activity have been hindered by the lack of suitable ligands for this enzyme. For instance, the metabolism rate of [11C]harmine in the plasma is very high, thus making a reliable quantification of the input function challenging 8 , and [11C]brofaromine has very little displaceable binding 9 .

Befloxatone is a potent, selective, and reversible inhibitor of brain MAO-A activity 10 . It can be radiolabeled with 11C using [11C]phosgene as the reagent 11 . [11C]Befloxatone has a highly specific binding to MAO-A, as shown by extensive in vitro 10 and in vivo studies in primates 12 . In particular, nonsaturable uptake, obtained after a pretreatment with a high dose of unlabeled befloxatone, is very low and represents only 3% of the total uptake 12 . [11C]Befloxatone has good imaging characteristics and has been used in positron emission tomography (PET) studies to quantify in vivo MAO-A in rats 13 , nonhuman primates 12 14 , and humans. In humans, it has been used to compare MAO-A binding potential in smokers and nonsmokers 7 ; however, a rigorous kinetic modeling study in humans has never been performed. This paper sought to (1) quantify the uptake of [11C]befloxatone in the human brain, both at the regional and voxel level, using an arterial input function; and (2) explore the possibility of replacing the arterial input function with a less invasive approach based on sampling of arterialized venous blood.

Methods

Subjects

Seven healthy, nonsmoker, male volunteers (72 ± 9.5 kg, age 26 ± 4.5 years) from a previous clinical protocol were included in this study 7 . All subjects were free of current medical or psychiatric illnesses and had no history of drug or alcohol abuse. Subjects were instructed to avoid monoamine-rich food 3 days before and alcohol consumption 1 week before imaging. Vital signs (blood pressure, heart rate) were recorded before [11C]befloxatone injection and at the end of the scan. The study was approved by the regional ethics committee for biomedical research at the Bicêtre Hospital. Each subject gave a written informed consent.

Brain image acquisition

PET scans were performed on an ECAT EXACT HR + scanner (Siemens Medical Solutions, Knoxville, TN, USA). One transmission scan was acquired before intravenous bolus injection of [11C]befloxatone (290 ± 24.4 MBq). The specific activity was 35.2 ± 14.9 GBq/μmol, which is within the range of the values usually obtained from the synthesis of [11C]befloxatone 11 and sufficient to rule out any mass effect. Even when considering a conservative B max value of 99 pmol/mL (as found in vitro in the human cerebellum 14 ), the B max occupancy is well below 1% in each subject. Dynamic three-dimensional images were acquired over 60 min in 29 frames, whose duration ranged from 10 s to 5 min. Images were reconstructed with filtered back projection.

Each subject also underwent brain T1-weighed three-dimensional magnetic resonance imaging (MRI) using a 1.5 T Signa scanner (GE Healthcare, Milwaukee, WI, USA). MRI images were acquired with a voxel size of 0.938 × 0.938 × 1.3 mm3 and a thickness of 1.3 mm in axial slices.

Measurement of [11C]befloxatone in arterial and venous plasma

Arterial blood samples were drawn from the radial artery of each subject in 10-s intervals until 2 min, followed by samples at 3, 4, and 5 min and then every 5 min until the end of the scan at 60 min. About 27 samples were drawn for each subject. For each blood sample, the decay-corrected concentration of [11C]befloxatone was measured in whole blood and plasma. Plasma was obtained by centrifugation. Plasma-free fraction was not measured. Venous blood samples were simultaneously drawn from the cubital vein of the opposite arm, using the same number of samples and time schedule as the arterial ones. As for arterial blood, plasma was separated from total venous blood. Arterialization of venous blood was achieved by heating the arm with hot water bags, starting 30 min before ligand injection until the end of the scan. The bags were replaced every 20 to 30 min. Four or five times during each scan, blood gases (pO2 and pCO2) were measured in five subjects.

The whole plasma time-activity curves were taken as input functions because in a previous series of (yet unpublished) four human scans from our laboratory, chromatograms obtained with high-performance liquid chromatography (HPLC) did not show any detectable radiometabolites in the plasma, even at late time points. The technique we used to measure the amount of unchanged radiotracer in the plasma was similar to that described in 14 . Specifically, after deproteinization with acetonitrile, the samples were centrifuged and the supernatant was injected directly into the HPLC column. A reverse-phase μBondapak C18 column (300 × 7.8 mm, 10 μm; Waters, Milford, MA, USA) was eluted applying a gradient from 20% acetonitrile in 0.01 M phosphoric acid up to 80% in 5.5 min, to 90% at 7.5 min, and returning to 20% at 7.6 min with a total run length of 10 min. The flow rate of the eluent was maintained at 6 mL/min. Befloxatone was eluted with a retention time of 6 min. Data acquisition and analysis were carried out using Winflow software (version 1.21, JMBS Developments, Grenoble, France).

Brain image analysis

The summed PET image, obtained by averaging all frames, was first coregistered to the individual MRI using SPM5 (Wellcome Department of Cognitive Neurology, London, UK). MRI and PET images were then normalized to the Montreal Neurologic Institute (MNI) space using the transformation parameters from the MRI images. A template of regions of interest (ROIs) 15 was used to extract brain time-activity curves for the following 13 regions: anterior and posterior cingulate cortices, caudate, putamen, thalamus, pons, cerebellum, hippocampus, parahippocampus, and the prefrontal, parietal, temporal, and occipital cortices. Each region was obtained from the weighted average of the left and right region. Image analysis and kinetic modeling were performed with Pmod 3.1 (Pmod Technologies, Zurich, Switzerland).

Calculation of distribution volume

Because [11C]befloxatone is a reversible MAO-A inhibitor 11 , kinetic analysis was performed using one- (1TCM) and two-tissue compartment (2TCM) model to calculate the total distribution volume (V T), which equals the ratio at equilibrium of the concentration of radioligand in the brain to that in the plasma 16 . V T is the sum of the binding in the specific compartment (V S) and in the nondisplaceable compartment (V ND). In theory, V T may change not only as a function of V S but also of V ND. However, in kinetic modeling studies, V ND is commonly assumed to be constant across different brain regions and across different subjects of the same species. Thus, variations of V T are considered to reflect variations of V S values. For [11C]befloxatone in particular, in vivo studies in primates showed that, after pharmacological blockade, the residual radioactive concentrations were very low and identical in all brain structures 11 . The 2TCM analysis was performed in two ways, (1) without constraining the rate constants and (2) fixing the ratio K 1/k 2 to the value obtained in the whole brain, to improve the identifiability of parameters. The individual rate constants were estimated with the weighted least square method and with the Marquardt optimizer. Brain activity was corrected for its vascular component using the measured whole blood concentrations, assuming that the cerebral blood volume is 5% of the total brain volume 17 . The delay of radiotracer arrival between the radial artery and the brain was corrected by fitting the whole brain. The plasma input function was modeled using a linear interpolation of the [11C]befloxatone concentrations before the peak and a tri-exponential fit of concentrations after the peak. Kinetic modeling was also performed at the regional level using the graphical Logan plot (Loganvoi), Ichise's multilinear analysis (MA1voi), and a standard spectral analysis (SAvoi).

Logan plot

The Logan plot is a model-independent graphical method for reversible tracers 18 . It performs a linearization of the data so that, after a certain time (t > t*), the slope can be related to the V T, according to

0 t C T τ C T t = V T 0 t C P τ C T t + c

where C T represents the tissue concentration and C p the plasma activity.

The Logan plot is a computationally quick and robust technique. It may, however, underestimate V T, especially in the case of small and noisy regions 19 .

Multilinear analysis

The multilinear analysis (MA1) is a modification of the Logan plot developed to remove noisy measurements (C T) from the independent variables 20 , according to the following equation:

C T t = - V b 0 t C p τ + 1 b 0 t C T τ

where V is the total distribution volume and b is the intercept of the Logan plot that becomes constant at t > t*. This approach has been shown to minimize V T estimation bias in the case of noisy measurements 20 .

Spectral analysis

This technique is based on a time-invariant single input/single output model used to identify tissue kinetic components 21 . As with the Logan plot, SA does not require prior knowledge of the number of compartments in the system. The tissue concentration, C T(t), is obtained by convolving the plasma time-activity curve, C P(t), with the sum of M + 1 exponentials, as follows:

C T t = j = 0 M C p t α j e - β j t = j = 0 M α j 0 t C p τ e - β j t - τ

where α j and β j are assumed to be real-valued and nonnegative.

V T was calculated from the estimated spectrum as

V T = j = 1 M α j β j

In the present study, the β j grid was defined with a maximum value of 100 and a logarithmic distribution of β j , j = 1, 2… M, using a spectral range from 0.01 to 1. SA was performed using the SAKE software 22 .

Finally, we sought to determine whether V T values could be obtained from voxel-wise analysis. We created parametric images using the Logan plot (Loganvoxel), multilinear analysis (MA1voxel), and standard spectral analysis (SAvoxel). For Loganvoxel and MA1voxel, the PET frames used for regression were selected on the basis of the start time obtained from the time-activity curve of the whole brain. For all parametric images, regional values were obtained by averaging the voxel values within each region.

Statistical analysis

Goodness of fit by nonlinear least squares analysis was evaluated using the Akaike information criterion (AIC) and model selection criterion (MSC). The most appropriate model is that with the smallest AIC and the largest MSC score. Goodness of fit by the one- and two-compartment models was compared with F statistics 23 . A value of P < 0.05 was considered significant. The identifiability of kinetic variables was expressed as a percentage of the rate constant and calculated as standard error obtained from the diagonal of the covariance matrix 24 . Standard error for V T was calculated from the covariance matrix using the generalized form of error propagation equation 25 , where correlations among parameters are taken into account. Smaller values indicate better identifiability.

For each subject, regional V T values obtained with 2TCM at the regional level were considered as the gold standard. Comparisons to the V T calculated with Loganvoi, MA1voi, SAvoi, Loganvoxel, MA1voxel, and SAvoxel were performed with a repeated measures analysis of variance (ANOVA) for V T values in the various regions. Statistical analyses were performed with SPSS (version 17 for Windows; SPSS Inc., Chicago, IL, USA).

Results

Measurement of [11C]befloxatone in arterial and venous plasma

[11C]Befloxatone concentrations in arterial plasma peaked about 30 s after injection at 21 ± 7.1 standard uptake value (SUV) and then rapidly decreased. Both the plasma and whole blood arterial curves were well fitted by a tri-exponential function. Venous concentrations peaked about 50 s after injection with an SUV of 6.7 ± 2.6.

Pairs of arterial and venous samples were sometimes drawn with a few seconds of delay in either direction. Therefore, in order to compare arteriovenous concentrations of the radioligand, we fitted the plasma time-activity curves with a tri-exponential function and interpolated the values at the same grid of standard times. The average pO2 measured in venous blood was 77% ± 8% of the pO2 measured in the arterial blood. The average pCO2 in the vein was 101% ± 1% compared to the arterial value. A stable arteriovenous equilibrium of [11C]befloxatone concentrations was found only a few minutes after injection (Figure 1). Average arteriovenous differences were already at about 5% or less within 2.5 min after injection and remained unchanged until the end of the scan (Figure 2). After the peak, in each individual subject the arteriovenous difference was almost always less than 10% at each time point. One subject, however, had differences between 10% and 20% at each time point from 40 min after injection to the end of the scan. A good correlation was found between arterial and venous concentrations at each time point from 3 to 60 min (n = 13 pairs). R 2 ranged from 0.64 at 3 min to 0.99 at 30 min.

<p>Figure 1</p>

Plasma time-activity curve of [11C]befloxatone concentrations in arterial and venous blood for a representative subject

Plasma time-activity curve of [11C]befloxatone concentrations in arterial and venous blood for a representative subject. The main frame shows the first 10 min; the inset magnifies the remaining part of the input function. Red curve, arterial blood; blue curve, venous blood. The venous peak was predictably much lower than the arterial one, but [11C]befloxatone concentrations were remarkably similar immediately after the peak until the end of the scan. Although compartmental modeling cannot be used with a venous input due to the different shapes of the early part of the curves, the Logan graphical analysis only depends on the total area under the curve. For this subject, the VT error using only the venous input was 7.2% compared to the reference VT values obtained with the arterial input.

<p>Figure 2</p>

Arteriovenous difference of radioligand concentrations (mean ± SD) across the different subjects

Arteriovenous difference of radioligand concentrations (mean ± SD) across the different subjects. Equilibrium was reached within 3 min of injection and remained stable until the end of the scan.

Brain radioactivity and kinetic analysis

Radioactivity showed a rapid uptake in the brain, progressively increased during the first 25 min (mean peak value, 1.5 to 2 SUV), and then slowly decreased (Figure 3). These time-activity curves were very similar to those described for [11C]befloxatone in nonhuman primates 12 .

<p>Figure 3</p>

Representative brain time-activity curves in one subject

Representative brain time-activity curves in one subject. Three regions are shown: putamen (empty circle), parietal cortex (filled circle), and cerebellum (empty square). Lines represent fitting with unconstrained 2TCM.

Radioactivity uptake was widespread and distributed fairly homogeneously across the brain (Figure 4). Consistent with the known distribution of MAO-A 2 , a lower uptake was noted in the cerebellum of all subjects (the cerebellar area under the curve was about 35% lower than the mean value of the whole brain). However, because low amounts of MAO-A are expressed in the cerebellum 26 , analysis with a reference region was not used.

<p>Figure 4</p>

Brain images

Brain images. MRI (A), summed PET (B), and VT parametric images in a representative subject. The parametric images were calculated with the Logan plot (C) and spectral analysis (D) and are shown with the same color scale. Parametric images obtained with MA1 (not shown) were too noisy to be quantified.

The brain kinetic analysis had two major results: (1) [11C]befloxatone binding in the brain was better described by a two-tissue compartment model but the 2TCM with a constraint of K 1/k 2 did not provide a better goodness of fit for V T, and (2) both at the regional and voxel level, SA provided better V T estimates than Logan and MA1.

As regards the first finding, unconstrained one- and two-compartment fitting converged in all regions and in all scans. Fitting was visually quite similar using either model; however, the two-tissue compartment model showed lower mean AIC (-22.3) and higher mean MSC (4.2) scores than the one-compartment model (AIC = 1.1, MSC = 3.4). The superiority of 2TCM was confirmed by F statistics. In addition, SA showed at least two equilibrating components in all regions for all subjects. The average identifiability across brain regions for two-compartment fitting was 4.7%. The average V T value across all brain regions was 14.3 ± 3.0 mL/cm3. The highest binding was found in the thalamus (average V T of 19.3 ± 1.4 mL/cm3), and the lowest in the cerebellum (6.7 ± 0.9 mL/cm3) (Table 1). The constrained two-compartment model fitting converged in all regions and in all scans but provided a similar goodness of fit (AIC = -20.8 and MSC = 4.1) as the unconstrained model. The average identifiability across brain regions was 4.4% (versus 4.7% for the unconstrained fitting).

<p>Table 1</p>

Region

V T (mL/cm 3 )

Region

Voxel

Unconstrained two-tissue compartment model

Logan

MA1

SA

Logan

SA

Values are presented as mean ± SD from seven human subjects. For each brain region, standard errors are listed in parentheses and are expressed as the percentage of the variable itself.

Frontal cortex

13.3 ± 0.84 (2.7)

12.5 ± 0.79

11.8 ± 0.71

13.6 ± 1.21

8.66 ± 2.14

13.7 ± 0.90

Temporal cortex

15.1 ± 1.04 (2.6)

14.1 ± 0.90

13.5 ± 0.83

15.3 ± 0.85

9.69 ± 2.64

15.4 ± 0.97

Hippocampus

16.2 ± 1.11 (6.9)

14.8 ± 1.09

13.8 ± 0.75

15.2 ± 1.74

6.36 ± 1.87

14.4 ± 1.79

Parietal cortex

13.0 ± 0.61 (2.8)

12.3 ± 0.53

11.6 ± 0.40

13.3 ± 0.94

8.59 ± 2.20

13.5 ± 1.10

Occipital cortex

15.4 ± 1.08 (2.8)

14.4 ± 0.99

13.6 ± 0.83

15.5 ± 1.20

9.87 ± 2.19

15.7 ± 1.13

Anterior cingulum

16.2 ± 1.56 (4.6)

15.2 ± 1.43

14.7 ± 1.20

16.9 ± 2.02

9.94 ± 2.56

16.2 ± 0.90

Caudate

12.5 ± 1.20 (7.7)

11.4 ± 1.00

11.0 ± 0.73

12.4 ± 1.21

7.17 ± 1.94

13.0 ± 1.36

Putamen

14.8 ± 0.67 (3.8)

13.9 ± 0.77

13.3 ± 0.70

15.2 ± 0.80

10.3 ± 3.00

16.5 ± 0.99

Thalamus

19.3 ± 1.43 (5.6)

17.7 ± 1.69

16.5 ± 1.18

19.5 ± 2.33

10.1 ± 3.11

19.0 ± 2.32

Cerebellum

6.73 ± 0.91 (2.5)

6.42 ± 0.82

6.18 ± 0.75

6.83 ± 0.95

5.19 ± 1.98

7.58 ± 0.86

Distribution volume calculated from compartmental modeling, Logan plot, MA1, and SA using regions or individual voxels

As regards the second point, Loganvoi and MA1voi underestimated V T values by about 10% (Figure 5A), and these differences were statistically significant (p < 0.05; factorial repeated measures ANOVA using Bonferroni adjustment, F 1.33,8.01 = 57.93). In contrast, V T values obtained with SAvoi had an average bias of only +1% (range -7% to +6%) and were not statistically different from the reference values obtained with 2TCM (p > 0.05). Loganvoi, MA1voi, and SAvoi correlated well with the reference V T values obtained with the 2TCM: R 2 was >0.90 for all subjects, except for one subject in whom R 2 for SAvoi was 0.73. Loganvoxel severely underestimated V T, with an average difference of -40% (p < 0.05) (Figure 5B). Moreover, linear regression analysis showed that Loganvoxel was significantly correlated to 2TCM V T values in only four of seven subjects. The parametric images obtained with MA1 (MA1voxel) contained many noisy voxels that prevented reliable quantification. These noisy voxels were located predominantly, but not exclusively, in areas where blood flow (K 1) is low, such as the white matter. Attempts to eliminate these noisy voxels by adjusting K 1 values were not successful. Therefore, MA1voxel was not considered for further analysis. Compared to 2TCM, V T values obtained with SAvoxel showed a mean bias of only +2% (range, -10% to +7%) and were not statistically different (p > 0.05) from the reference values. Predictably, SAvoxel values were higher than Loganvoi, Loganvoxel, and MA1voi (p < 0.05). Finally, SAvoxel significantly correlated with 2TCM in all subjects: R 2 for brain regions ranged from 0.81 to 0.95, except in one subject who had an R 2 of 0.60.

<p>Figure 5</p>

Regression analyses for VT results obtained from the 13 brain regions of a 23-year-old subject

Regression analyses for VT results obtained from the 13 brain regions of a 23-year-old subject. (A) Spectral analysis (filled square) provided good results with a very small bias (+1.4%). Logan (empty circle) and MA1 (filled circle) slightly underestimated VT values (-7.4% and -9.7%, respectively, for this subject). At the voxel level (B), only SA provided results with a small bias and good correlation with 2TCM (0.932x + 1.793, R2 = 0.92, p < 0.0001). In contrast, Loganvoxel underestimated VT values by about 47% (0.324x + 2.61, R2 = 0.32, p = 0.047).

Use of a venous input function

Logan-V T results obtained with a full venous input function were consistent with Logan-V T results obtained with an arterial plasma input function in six subjects (with an error <10% compared to the reference values obtained with the arterial input function). However, in one subject, Logan-V T was overestimated by 27%.

Discussion

This study sought to quantify [11C]befloxatone binding in human brain using arterial blood samples and PET imaging. We also assessed whether arterial cannulation could be avoided by replacing arterial samples with arterialized venous samples.

We found that [11C]befloxatone binding is better quantified using a two-tissue compartment model or SA. Venous samples cannot, however, be a reliable substitute for arterial samples.

Despite a slow washout of the radioligand from the brain, V T values were well identified with an unconstrained two-tissue compartment model (mean standard error, 4.7%). A constrained 2TCM did not provide a clear advantage for V T estimation over the unconstrained model. V T values were higher in the basal ganglia and the cortical structures (about 12 to 19 mL/cm3) and lower in the cerebellum (about 7 mL/cm3). These values were slightly lower than the V T values previously found in baboons, where the V T was 19 to 23 mL/cm3 for high-uptake regions and 10 to 11 mL/cm3 for the cerebellum 12 . Notably, in a previous study in baboons, we used a three-compartment model, corresponding to the free ligand in the tissue, the ligand specifically bound to the receptors and the nonspecific binding 14 . However, the concentration of the nonspecific binding was very low in all regions, and the equilibrium between the free and the nonspecific ligand was reached relatively quickly, compared with the equilibrium between the other compartments 14 . Therefore, the nonspecific and the free compartments of [11C]befloxatone can be lumped together in a two-tissue model.

Compared with 2TCM, both the Logan plot (Loganvoi) and the multilinear analysis (MA1voi) significantly underestimated V T values in large brain regions (about -10%). In contrast, the average bias of V T values obtained with standard SAvoi was only 1%.

Owing to the sensitivity of the Logan plot to noise, especially at the voxel level, Loganvoxel severely underestimated V T values (-40% on average) and did not significantly correlate with 2TCM V T in three out of the seven subjects. Multilinear analysis has been shown to be more robust to noise-induced bias 20 . However, parametric images (MA1voxel) contained numerous noisy voxels and were therefore not amenable to quantification. SA was the only technique capable of providing good results at the voxel level. Indeed, SAvoxel showed only a small, nonsignificant bias (+2%) compared to V T values obtained with 2TCM and correlated well with 2TCM for all subjects. As this study shows, when the most commonly used graphical approaches fail to provide parametric images of good quality, SA can be a valid alternative. The use of SA has long been hampered by the complexity of its implementation. However, new user-friendly software, such as SAKE or Pmod 3.4, may encourage its use for clinical protocols where statistical parametric mapping analyses are performed.

Autoradiography and immunohistochemistry data have demonstrated that MAO-A is widespread in the brain 27 28 . Moreover, studies of nonhuman primates found that [11C]befloxatone binding was displaceable in all brain regions 12 . These data suggest that a reference region may not exist to quantify [11C]befloxatone binding. Without a reference region, quantification can only be achieved by using an arterial plasma input function, which is an invasive procedure. In addition, measuring the arterial input function usually requires labor-intensive HPLC analyses. However, one of the favorable characteristics of [11C]befloxatone is that no radiometabolites are detected in the plasma. In consequence, [11C]befloxatone input function can be easily achieved using only plasma isolation without HPLC analyses. Notably, our previous analysis in baboons showed that [11C]befloxatone radiometabolites constituted about 20% of plasma radioactivity at 30 min after injection 14 . This discrepancy was due to interspecies differences and possibly linked to the faster metabolism of baboons.

To avoid arterial samples altogether, we sought to validate the use of an input function composed of arterialized venous samples. Indeed, arterialized venous samples are routinely used in several institutions as a replacement for arterial samples. A stable arteriovenous equilibrium began from about 3 min after injection; therefore, the early portions of the arterial and venous input functions (i.e., the peaks) were dissimilar. This is a common finding with most tracers and molecules and is linked to the diffusion of the radiotracer in its volume of distribution 29 30 . A different peak shape would prevent the use of a venous full input in association with compartmental modeling or SA. In contrast, the Logan plot does not require that the shape of the peak be perfectly estimated because it relies on the total area under the curve of the input function. Indeed, kinetic modeling performed using a venous input function yielded Logan-V T values which were consistent with those obtained from arterial sampling: six out of seven subjects had Logan-V T estimation errors smaller than 10%. There was, however, one major outlier with an error of 27%, and this suggests that arterialized samples might not be a reliable alternative to the full arterial input function for [11C]befloxatone imaging. Notably, this subject was the only one to have a significant deviation from arteriovenous equilibrium at a late time. However, the degree of arterialization of the venous blood for this subject (vein/artery ratio of 0.74 for pO2 and 1.03 for pCO2) was similar to that of the others (0.77 ± 0.08 and 1.01 ± 0.01 for pO2 and pCO2, respectively). It is possible that a higher degree of arterialization could have been achieved using a different warming technique 31 . It should, however, be noted that, even with optimal conditions, arterialization of venous blood is difficult to achieve and is plagued by intersubject variability 32 . Therefore, unless a better equilibrium can be achieved with a different technique, an arterial input function will be necessary for the quantification of [11C]befloxatone binding.

Finally, it should be noted that despite having several good imaging characteristics (high specific binding, low metabolism in plasma), [11C]befloxatone has one important drawback, i.e., a slow washout from the brain. Although V T values were well identified, it would nevertheless be preferable to have a radioligand with a faster washout, even if there was a higher rate of metabolism in the plasma (provided that radiometabolites do not enter into the brain).

Conclusions

Binding of [11C]befloxatone to MAO-A can be quantified using an arterial input function and a two-compartment model or, in parametric images, with spectral analysis.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

PZF conceived of the study, performed the image analyses and statistical comparisons, and drafted the manuscript. CL, DR, CT, JLM, and MB recruited the subjects, acquired the image and blood data and participated in the data analysis and the design, coordination, and writing of the manuscript. All authors read and approved the final manuscript.

Acknowledgements

The authors are grateful to Gaia Rizzo and Mattia Veronese for their help in implementing the spectral analysis and to Masahiro Fujita for the useful comments on kinetic modeling. Ioline Henter provided excellent editorial assistance.

<p>Inhibition of monoamine oxidase B in the brains of smokers</p>FowlerJSVolkowNDWangGJPappasNLoganJMacGregorRAlexoffDSheaCSchlyerDWolfAPWarnerDZezulkovaICilentoRNature199637973373610.1038/379733a08602220<p>Monoamine oxidase and tobacco dependence</p>LewisAMillerJHLeaRANeurotoxicology20072818219510.1016/j.neuro.2006.05.01916859748<p>Monoamine oxidases in major depressive disorder and alcoholism</p>DuncanJJohnsonSOuXMDrug Discov Ther2012611212222890201<p>Monoamine oxidase inhibition dramatically increases the motivation to self-administer nicotine in rats</p>GuillemKVouillacCAzarMRParsonsLHKoobGFCadorMStinusLJ Neurosci2005258593860010.1523/JNEUROSCI.2139-05.200516177026<p>Monoamine oxidase inhibitors allow locomotor and rewarding responses to nicotine</p>VillegierASSalomonLGranonSChangeuxJPBelluzziJDLeslieFMTassinJPNeuropsychopharmacology2006311704171310.1038/sj.npp.130098716395299<p>Brain monoamine oxidase A inhibition in cigarette smokers</p>FowlerJSVolkowNDWangGJPappasNLoganJSheaCAlexoffDMacGregorRRSchlyerDJZezulkovaIWolfAPProc Natl Acad Sci USA199693140651406910.1073/pnas.93.24.14065194958943061<p>Cerebral monoamine oxidase A inhibition in tobacco smokers confirmed with PET and [<sup>11</sup>C]befloxatone</p>LeroyCBragulatVBerlinIGregoireMCBottlaenderMRoumenovDDolléFBourgeoisSPenttiläJArtigesEMartinotJLTrichardCJ Clin Psychopharmacol200929868810.1097/JCP.0b013e31819e98f19142115<p>11C-harmine as a tracer for monoamine oxidase A (MAO-A): in vitro and in vivo studies</p>BergstromMWesterbergGLangstromBNucl Med Biol19972428729310.1016/S0969-8051(97)00013-99257326<p>Radiosynthesis of [<sup>11</sup>C]brofaromine, a potential tracer for imaging monoamine oxidase A</p>AmetameySMBeerHFGuentherIAntoniniALeendersKLWaldmeierPCSchubigerPANucl Med Biol19962322923410.1016/0969-8051(95)02051-98782230<p>Befloxatone, a new reversible and selective monoamine oxidase-A inhibitor</p>CuretODamoiseauGAubinNSontagNRoveiVJarreauFXI Biochem profile J Pharmacol Exp Ther1996277253264<p>Synthesis and in vivo imaging properties of [<sup>11</sup>C]befloxatone: a novel highly potent positron emission tomography ligand for mono-amine oxidase-A</p>DolleFValetteHBramoulleYGuentherIFuseauCCoulonCLartizienCJeghamSGeorgePCuretOPinquierJLBottlaenderMBioorg Med Chem Lett2003131771177510.1016/S0960-894X(03)00215-412729662<p>Mapping the cerebral monoamine oxidase type A: positron emission tomography characterization of the reversible selective inhibitor [<sup>11</sup>C]befloxatone</p>BottlaenderMDolleFGuentherIRoumenovDFuseauCBramoulleYCuretOJeghamJPinquierJLGeorgePValetteHJ Pharmacol Exp Ther200330546747310.1124/jpet.102.04695312606609<p>Acute inhibition of cardiac monoamine oxidase A after tobacco smoke inhalation: validation study of [<sup>11</sup>C]befloxatone in rats followed by a positron emission tomography application in baboons</p>ValetteHBottlaenderMDolleFCoulonCOttavianiMSyrotaAJ Pharmacol Exp Ther200531443143610.1124/jpet.105.08570415833896<p>In vivo quantification of monoamine oxidase A in baboon brain: a PET study using [(11)C]befloxatone and the multi-injection approach</p>BottlaenderMValetteHDelforgeJSabaWGuentherICuretOGeorgePDolléFGrégoireMCJ Cereb Blood Flow Metab20103079280010.1038/jcbfm.2009.242294915919920845<p>Automated anatomical labeling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single-subject brain</p>Tzourio-MazoyerNLandeauBPapathanassiouDCrivelloFEtardODelcroixNMazoyerBJoliotMNeuroimage20021527328910.1006/nimg.2001.097811771995<p>Consensus nomenclature for <it>in vivo</it> imaging of reversibly binding radioligands</p>InnisRBCunninghamVJDelforgeJFujitaMGjeddeAGunnRNHoldenJHouleSHuangSCIchiseMIidaHItoHKimuraYKoeppeRAKnudsenGMKnuutiJLammertsmaAALaruelleMLoganJMaguireRPMintunMAMorrisEDParseyRPriceJCSlifsteinMSossiVSuharaTVotawJRWongDFJ Cereb Blood Flow Metab2007271533153910.1038/sj.jcbfm.960049317519979<p>Cerebral blood flow and glucose metabolism in healthy volunteers measured using a high-resolution PET scanner</p>HuismanMCvan GolenLWHoetjesNJGreuterHNSchoberPIjzermanRGDiamantMLammertsmaAAEJNMMI Res201226310.1186/2191-219X-2-63354465323168248<p>Graphical analysis of reversible radioligand binding from time-activity measurements applied to [N-<sup>11</sup>C-methyl]-(-)cocaine PET studies in human subjects</p>LoganJFowlerJSVolkowNDWolfAPDeweySLSchlyerDJMacGregorRRHitzemannRBendriemBGatleySJChristmanDRJ Cereb Blood Flow Metab19901074074710.1038/jcbfm.1990.1272384545<p>Effects of statistical noise on graphic analysis of PET neuroreceptor studies</p>SlifsteinMLaruelleMJ Nucl Med2000412083208811138696<p>Strategies to improve neuroreceptor parameter estimation by linear regression analysis</p>IchiseMToyamaHInnisRBCarsonREJ Cereb Blood Flow Metab2002221271128112368666<p>Spectral analysis of dynamic pet studies</p>CunninghamVJJonesTJ Cereb Blood Flow Metab199313152310.1038/jcbfm.1993.58417003<p>SAKE: a new quantification tool for positron emission tomography studies</p>VeroneseMRizzoGTurkheimerFEBertoldoAComput Methods Programs Biomed201311119921310.1016/j.cmpb.2013.03.01623611334<p>Effects of temporal sampling, glucose metabolic rates, and disruptions of the blood–brain barrier on the FDG model with and without a vascular compartment: studies in human brain tumors with PET</p>HawkinsRAPhelpsMEHuangS-CJ Cereb Blood Flow Metab1986617018310.1038/jcbfm.1986.303485641<p>Parameter estimation in positron emission tomography</p>CarsonREPositron Emission Tomography and Autoradiography: Principles and Applications for the Brain and HeartNew York: Raven PressPhelps ME, Mazziotta JC, Schelbert HR1986347390BevingtonPRRobinsonDKData Reduction and Error Analysis for the Physical SciencesNew York: McGraw-Hill2003<p>Calcium alters monoamine oxidase-A parameters in human cerebellar and rat glial C6 cell extracts: possible influence by distinct signalling pathways</p>CaoXLiXMMousseauDDLife Sci20098526226810.1016/j.lfs.2009.06.00419539632<p>Molecular neuroanatomy of monoamine oxidases in human brainstem</p>RichardsJGSauraJUlrichJDa PradaMPsychopharmacology (Berl)1992106SupplS21S231546134<p>Molecular neuroanatomy of human monoamine oxidases A and B revealed by quantitative enzyme radioautography and in situ hybridization histochemistry</p>SauraJBleuelZUlrichJMendelowitschAChenKShihJCMalherbePDa PradaMRichardsJGNeurosci19967075577410.1016/S0306-4522(96)83013-2<p>The phenomenon and rationale of marked dependence of drug concentration on blood sampling site. Implications in pharmacokinetics, pharmacodynamics, toxicology and therapeutics (Part II)</p>ChiouWLClin Pharmacokinet19891727529010.2165/00003088-198917040-000052686884<p>The phenomenon and rationale of marked dependence of drug concentration on blood sampling site. Implications in pharmacokinetics, pharmacodynamics, toxicology and therapeutics (Part I)</p>ChiouWLClin Pharmacokinet19891717519910.2165/00003088-198917030-000042680213<p>Methodological considerations in arterialization of venous blood</p>FraynKNMacdonaldIAClin Chem1992383163171541023<p>Invalidity of hand heating as a method to arterialize venous blood</p>GreenJHEllisFRShallcrossTMBramleyPNClin Chem1990367197222337978