Acquisition setting optimization and quantitative imaging for 124I studies with the Inveon microPET-CT system. - Inserm - Institut national de la santé et de la recherche médicale Accéder directement au contenu
Article Dans Une Revue EJNMMI Research Année : 2012

Acquisition setting optimization and quantitative imaging for 124I studies with the Inveon microPET-CT system.

Résumé

ABSTRACT: BACKGROUND: Noninvasive multimodality imaging is essential for preclinical evaluation of the biodistribution and pharmacokinetics of radionuclide therapy and for monitoring tumor response. Imaging with nonstandard positron-emission tomography [PET] isotopes such as 124I is promising in that context but requires accurate activity quantification. The decay scheme of 124I implies an optimization of both acquisition settings and correction processing. The PET scanner investigated in this study was the Inveon PET/CT system dedicated to small animal imaging. METHODS: The noise equivalent count rate [NECR], the scatter fraction [SF], and the gamma-prompt fraction [GF] were used to determine the best acquisition parameters for mouse- and rat-sized phantoms filled with 124I. An image-quality phantom as specified by the National Electrical Manufacturers Association NU 4-2008 protocol was acquired and reconstructed with two-dimensional filtered back projection, 2D ordered-subset expectation maximization [2DOSEM], and 3DOSEM with maximum a posteriori [3DOSEM/MAP] algorithms, with and without attenuation correction, scatter correction, and gamma-prompt correction (weighted uniform distribution subtraction). RESULTS: Optimal energy windows were established for the rat phantom (390 to 550 keV) and the mouse phantom (400 to 590 keV) by combining the NECR, SF, and GF results. The coincidence time window had no significant impact regarding the NECR curve variation. Activity concentration of 124I measured in the uniform region of an image-quality phantom was underestimated by 9.9% for the 3DOSEM/MAP algorithm with attenuation and scatter corrections, and by 23% with the gamma-prompt correction. Attenuation, scatter, and gamma-prompt corrections decreased the residual signal in the cold insert. CONCLUSIONS: The optimal energy windows were chosen with the NECR, SF, and GF evaluation. Nevertheless, an image quality and an activity quantification assessment were required to establish the most suitable reconstruction algorithm and corrections for 124I small animal imaging.
Fichier principal
Vignette du fichier
2191-219X-2-7.pdf (540.39 Ko) Télécharger le fichier
2191-219X-2-7.xml (70.37 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Format : Autre
Loading...

Dates et versions

inserm-00682229 , version 1 (23-03-2012)

Identifiants

Citer

Nadège Anizan, Thomas Carlier, Cecilia Hindorf, Jacques Barbet, Manuel Bardiès. Acquisition setting optimization and quantitative imaging for 124I studies with the Inveon microPET-CT system.. EJNMMI Research, 2012, 2 (1), pp.7. ⟨10.1186/2191-219X-2-7⟩. ⟨inserm-00682229⟩
138 Consultations
420 Téléchargements

Altmetric