1532-429X-13-S1-O301532-429X Oral presentation <p>Evaluation of aortic valve stenosis from Phase-Contrast Magnetic Resonance data using a new automated segmentation and analysis method: Comparison against Doppler Echocardiography</p> DefranceCarine BollacheEmilie KachenouraNadjia BruguièreEric RedheuilAlban DieboldBenoit PerdrixLudivine MousseauxElie

Hopital Européen Georges Pompidou, Paris, France

INSERM U678, Paris, France

Journal of Cardiovascular Magnetic Resonance <p>Abstracts of the 2011 SCMR/Euro CMR Joint Scientific Sessions</p> Meeting abstracts - A single PDF containing all abstracts in this supplement is available here. <p>2011 SCMR/Euro CMR Joint Scientific Sessions</p> Nice, France 3-6 February 2011 http://www.scmr.org/ 1532-429X 2011 13 Suppl 1 O30 http://jcmr-online.com/content/13/S1/O30 10.1186/1532-429X-13-S1-O30
222011 2011Defrance et al; licensee BioMed Central Ltd.This is an open access article distributed under the terms of the Creative Commons Attribution License (<url>http://creativecommons.org/licenses/by/2.0</url>), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Background

Aortic valve stenosis (AVS) is the most common valvular disease. Its evaluation is of growing interest because of its increasing incidence with the aging population. Previous studies demonstrated the usefulness of Phase-Contrast Magnetic Resonance (PCMR) images in the evaluation of AVS. However, because of the lack of automated methods for PCMR data analysis, this technique remains time-consuming and operator-dependent.

Objectives

Therefore, the aims of this study were 1) to develop a semi-automated method for aortic flow analysis from PCMR images, and 2) to evaluate several approaches of aortic valve area (AVA) estimation.

Methods

We studied 37 consecutive patients with AVS (mean AVA:0,89 +/-0,42 cm2) and 12 healthy subjects (mean AVA: 3,19 +/- 0,65 cm2) who had the same day a trans-thoracic echocardiography (TTE) and PCMR acquisitions at the levels of the aortic valve and the left ventricle outflow tract (LVOT). PCMR data analysis included a semi-automated segmentation, based on pixels connectivity in terms of velocity sign, to delineate the aortic flow on all systolic frames, as well as a functional parameters extraction from aortic velocity and flow rate curves such as aortic maximal velocity (VmaxAo) and AVA. AVA was calculated using: 1) Hakki’s formula which is a simplification of Gorlin’s formula, resulting in AVA1= cardiac output divided by √systolic pressure gradient, 2) the continuity equation with the most pertinent method found in previous studies in PCMR resulting in AVA2 = LVOT stroke volume (LVOT SV) divided by aortic valve velocity time integral (VTIAo), and 3) the continuity equation with another approach previously described in echocardiography but never used in PCMR. It resulted in AVA3= LVOT peak Flow rate (QLVOT) divided by VmaxAo (figure 1).

<p>Figure 1</p>

A) Example of left ventricular outflow tract (LVOT) segmentation on a velocity PC MR image, flow rate curve, whose integral corresponds to LVOT stroke volume (LVOT SV) and whose peak corresponds to Q LVOT Q; B) Example of aortic flow segmentation on a velocity PC MR image, transaortic maximal velocity curve, whose integral corresponds to VTIAO and whose peak corresponds to VmaxAO. AVA2 is the ratio between the two integrals. AVA3 is the ratio between the two peaks.

A) Example of left ventricular outflow tract (LVOT) segmentation on a velocity PC MR image, flow rate curve, whose integral corresponds to LVOT stroke volume (LVOT SV) and whose peak corresponds to Q LVOT Q; B) Example of aortic flow segmentation on a velocity PC MR image, transaortic maximal velocity curve, whose integral corresponds to VTIAO and whose peak corresponds to VmaxAO. AVA2 is the ratio between the two integrals. AVA3 is the ratio between the two peaks.

Results

The LVOT and aortic flows were successfully delineated on all phases for all subjects. Comparison of PCMR and echocardiographic VmaxAo resulted in a good correlation(r = 0.92). Hakki’s formula underestimated the AVA with regard to TTE in the absence of severe stenosis. AVA3 provided the best results (Table 1) in terms of detection of severe stenosis (r = 0.97; Specificity = 100% and Sensibility = 97%). Moreover, excellent intra-observer reproducibility was found (AVA3: ICC > 0.99 and mean differences: 0.00 ± 0.02 cm2; VmaxAo: ICC > 0.99 and mean differences: 0.02 ± 0.10 m/s).

<p>Table 1</p>

Comparison between TTE and PC MRI using the three methods for AVA estimation (*p<0.05)

Mean (SD) of differences compared to TTE (cm2)

Correlation Coefficient R

Intra-class coefficient between TTE and PC MR

Specificity to detect severe stenosis (%)

Sensibility to detect severe stenosis (%)


AVA1

-0,50 (0,57)*

0,92

0 ,67

94

97

AVA2

+0,02 (0,37)

0,94

0,94

100

97

AVA3

+0,07 (0,29)

0,97

0,96

100

97

Conclusion

Our semi-automated approach for AVS evaluation from PCMR provided reproducible velocity measurements and AVA estimates in good agreement with echocardiographic values, and were able to characterize the severity of AVS.