New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices - Inserm - Institut national de la santé et de la recherche médicale Accéder directement au contenu
Article Dans Une Revue Computational and Mathematical Methods in Medicine Année : 2014

New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices

Résumé

Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5%, corresponding to an improvement of 14% compared to the standard estimator.
Fichier principal
Vignette du fichier
Voicu_12757.pdf (365.85 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-00881484 , version 1 (21-10-2014)
hal-00881484 , version 2 (04-03-2015)

Identifiants

Citer

Iulian Voicu, Sébastien Ménigot, Denis Kouamé, Jean-Marc Girault. New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices. Computational and Mathematical Methods in Medicine, 2014, vol. 2014 (Article ID 784862), pp. 1-10. ⟨10.1155/2014/784862⟩. ⟨hal-00881484v2⟩
241 Consultations
298 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More